 John V. Hinshaw
|
Gas chromatography (GC) inlets transform a sample from its physical state outside the chromatograph into a state suitable
for separation inside the column. This transformation consists of two principal steps: transfer from the outside into the
inlet and transfer from the inlet into the column. On the way to the column, the sample undergoes volume, concentration, temperature,
and pressure changes that convert it into a condition that is more or less compatible with the separation that follows. Chromatographers
measure an injection's success by the degree to which it preserves the relative sample composition while not interfering with
separation, and by how well the injection process repeats from run to run.
In GC, samples can be liquids, solids, or gases. The type of sample and its concentration, taken along with the column and
detector, determine which types of injection will give the best results. There often is more than one choice of appropriate
injection technique, of which only one might be available in a particular laboratory or instrument. Sometimes none of the
suitable inlet systems is available, in which case the analyst might have to choose between compromising the injection or
upgrading their chromatographic equipment. Where they have the methodological latitude, they also can solve this dilemma by
modifying preparation procedures to make the sample compatible with available inlet systems, or by choosing a different, more
compatible column. For example, analysis of a trace-level sample might best be performed with splitless or on-column injection
onto a 25-m, 0.25-mm i.d. high-resolution column. Without an on-column or split–splitless inlet on-hand, the chromatographer
could choose to sacrifice speed of analysis by using a longer 60-m, 0.53-mm i.d. megabore column with the same phase ratio
and direct injection. Such a column could produce about the same resolution as the narrow-bore column in about twice the time,
and it is compatible with the larger injection volumes produced by direct injection.
Liquid sample injection techniques for open-tubular (capillary) column GC can be classified by, among several variables, the
fraction of the vaporized sample that enters the column. For trace-level analyses, it is desirable for all or at least a substantial
portion of the sample to enter the column to maximize method sensitivity. But at higher concentrations the amounts of analytes
that traverse the column can become large enough to affect separation quality by causing peak overloading and peak shape distortion,
so in these cases only a fraction of the sample is passed into the column by using split injection. To complicate the situation,
passing liquid sample volumes greater than about 1 μL directly onto the column can cause a number of solvent-induced side
effects that detract from the column's resolving power: controlling these effects can be critical. Often, it is very clear
which type of injection technique will be the best choice for a specific sample and analysis, but there are many samples that
could be handled by multiple injection types. Deciding which to choose can be difficult, and understanding the side effects
that can occur is vital for troubleshooting inlet problems.